Fentanyl Test Strips
The Rapid Response Fentanyl Test Strip is an easy-to-use tool to detect fentanyl in liquid or powder drug samples.
Testing Procedure
-
Place a small amount (5-10 mg) of the drug into a clean container.
-
Add 5 mL of water and mix until dissolved.
-
Hold the strip by the blue end, where the product name (FYL) is printed. Do not touch the white section of the strip
-
Holding the strip vertically, dip the wavy end of the test strip into the liquid for 15 seconds. Do not dip past the solid maximum line.
-
Remove the strip. Place it on a flat, non-absorbent surface for 5 minutes.
-
Read the results. Do not read the results after 10 minutes.
Results Interpretation
Positive- Fentanyl DetectedOnly one colored line appears in the control region (C). No apparent line appears in the test region (T). |
![]() |
Negative- Fentanyl Could Not be DetectedTwo colored lines appear on the test strip. One line appears in the control region (C) and another line appears in the test region (T). |
|
InvalidControl line does not appear. If a control line does not appear after 5 minutes, the test must be thrown away. Review the testing procedure and repeat with a new test. |
|
How to Use the Rapid Response Fentanyl Test Strip